1
Introduction
Thischapterprovidesgeneralinformationrequiredfortheproperuseoftheproduct
andthemanual.Familiarizeyourselfwiththisinformationbeforeusingtheproduct.
ProductInformation
Thissectionprovidestheindicationsforuse,generaldescription,andidentication
informationforboththeCAM-14andCAM-HDacquisitionmodules.Exceptwhere
differencesbetweenthetwoproductsarenoted,thetermsacquisitionmoduleand
devicerefertobothproducts.
IndicationsforUse
Theacquisitionmoduleisintendedtoacquire,digitize,andtransmitanalogECG
signalstoahostunit.Thecircuitryisdesignedtoprotectthehostunitagainstthe
effectsofcardiacdebrillatordischargetoensuredevicerecovery.
Theacquisitionmoduleisintendedtobeusedunderthedirectsupervisionofa
licensedhealthcarepractitioner.
GeneralDescription
TheacquisitionmoduleperformshighresolutionECGdataacquisitionforusewith
hostequipment(restingECGanalysissystemsandexercisesystems).Theacquisition
modulehasthefollowingfeatures:
•ACleadfailbias
•leadoffdetection
•patientisolation
•functionkeycontrolofhostequipmentfunctions
Theacquisitionmoduleprovidespatientelectricalisolationforthehostequipment.
TheminuteECGsignalsfromthepatient’sskinarereceivedbytheelectrodesand
senttotheacquisitionmodulevialeadwires.Theacquisitionmodulethenamplies,
digitizes,andperformsprocessingonthesignals.
2049751-001C
CAM-14
®
/CAM-HD
®
5