EasyManua.ls Logo

GE CAM-14 - Electrical Safety Checks

GE CAM-14
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
ElectricalSafetyChecks
1.Conductcurrentleakageandgroundcontinuitytests.
Performelectricalsafetycheckswhenindicatedintheprecedingtable.All
indicatedelectricalsafetychecksrequireapass/failindicationforthesteps
performed.Recordthemeasurementvaluesinyourdebrief.Referto“Leakage
Tests”onpage19foradditionalinformation.
2.Conductthedielectricwithstandtest.
Referto“DielectricWithstandTest”onpage21foradditionalinformation.
LeakageTests
Theleakagetestsaresafetyteststoensurethattheequipmentposesnoelectrical
healthhazards.Usethefollowingtabletodeterminewhichtestsapplytotheunit
undertestandthemaximumallowableleakagecurrents.Forinternationalleakage
limits,refertotheinternalstandardsagenciesofthatparticularcountry.
Iftheunitundertestfailstheleakagetests,donotallowthecustomertousethe
equipment.CallGEHealthcareCustomerSupportforassistance.
GEHealthcarerecommendsthatyouperformthesetestswiththefollowingfrequency:
Beforeapplyingpowerforthersttime
Everyyearaspartofroutinemaintenance
Wheneverinternalassembliesareserviced
Youneedaleakagetestertoperformtheleakagetests.
NOTE:
Theaccuracyoftheleakagetestsdependsonaproperly-wiredwalloutlet.Donot
proceeduntilyouverifytheintegrityofthepowersource.
WARNING:
Totalsystemleakagecurrentmustnotexceed100microamperes.
2049751-001C
CAM-14
®
/CAM-HD
®
19