Loading...
Maintenance
ElectricalSafetyChecks
Step
Condition
1
UUTON
2
ResultLeakage
Current
Limits
PatientLeakageCurrenttoGround
1.
Forward
Polarity
NC
______µA
Pass/Fail
10µA
2.
Neutralopen,
Forward
Polarity
SFC
______µA
Pass/Fail
50µA
3.
Groundopen,
Forward
Polarity
SFC
______µA
Pass/Fail
50µA
4.
Groundopen,
ReversePolarity
SFC
______µA
Pass/Fail
50µA
5.
Neutralopen,
ReversePolarity
SFC
______µA
Pass/Fail
50µA
6.
ReversePolarity
NC
______µA
Pass/Fail
10µA
PatientLeakageCurrentMainsonAppliedPart
3
1.
Forward
PolarityNeutral
/GroundClosed
SFC
______µA
Pass/Fail
5000µA
2.
ReversePolarity
Neutral/
GroundClosed
SFC
______µA
Pass/Fail
5000µA
1.NC=NormalConditionSFC=SingleFaultConditionN/A=NotApplicable
2.UUT=UnitUnderTest
3.AllSIPs/SOPsgrounded
20