Troubleshooting
DisassemblingtheAcquisitionModule
1.RemovethefourTorxscrews(1)holdingthetopandbottomofthedevice
togetherusinga#6Torxdriver.
2.Liftthetopcover(2)highenoughtoallowaccesstotheswitchexcable(3)from
theswitchassembly.
3.Toavoiddamagetotheexcable(3),gentlyprythecableofftheswitchatthe
mainPCBconnector(4)withasmallscrewdriverbeforeremovingthecover.
NOTE:
Damagetotheswitchexcableconnectorwillresultinfailureofoneor
moreoftheswitchfunctions.
DONOTattempttodisconnecttheswitchexcablebypullingonit.
DONOTletthemainPCBhangbythecable.
4.WhenthemainPCBhasbeenseparatedfromtheswitchexcable,disconnect
thehostcableinterfaceex(5)fromthemainPCBbygentlypryingthe
connectorsapart.