Maintenance
FunctionalCheckoutProcedures
Thefollowingtableidentiesthetoolsandproceduresrequiredtoperformafunctional
checkoutafterreplacingthespeciedFRUorperformingthespeciedtask.Tousethe
table,locatetherelevantFRUortaskintherstcolumnandnotetherequiredTools,
VisualInspections,OperationalCheck,andElectricalSafetyCheck.Then,locatethe
referencedinstructionsinthecorrespondingsectionsthatfollowthetable.
NOTE:
Theeldreplaceableunit(FRU)checkoutprocedureforanylistedFRUalsoapplies
toitsinternalPCBsandcomponents.
Inadditiontothefollowingprocedures,youshouldalsoperformanycheckout
proceduresrequiredbythehostsystem.Refertothehostsystem'seldservice
manualforthehostsystem'scheckoutprocedures.
FunctionalCheckoutRequirements
FRUDescription
ToolsVisual
Inspection
Operational
Check
Electrical
SafetyCheck
FRURepairs
CAM-141,61,2,31,2,3
notapplicable
CAM-HD1,61,2,31,2,3
notapplicable
Allinternal
FRU's/parts,
Covers,and
fastener
replacements
1,2,3,4,5,61,2,31,2,31,2
Non-FRURepairs
Noparts
replaced
1,5,61,2,31,2,3
notapplicable
Annualelectrical
safetycheckout
PerHostDeviceRequirement
2049751-001C
CAM-14
®
/CAM-HD
®
17