CARESCAPE Monitor B850
110
2106778-001
NOTE: *The measuring device (MD) represents the network and voltage measuring instrument
and its frequency characters according to IEC 60601-1.
1. Configure the safety analyzer as follows (NC):
Polarity: NORMAL
Neutral: CLOSED
Earth (GND): CLOSED
2. Power on the device under test.
3. Read and record the current leakage indicated on the safety tester.
4. Configure the safety analyzer as follows (SFC):
Polarity: NORMAL
Neutral: OPEN
Earth (GND): CLOSED
5. Read and record the current leakage indicated on the safety tester.
6. Configure the safety analyzer as follows (SFC):
Polarity: NORMAL
Neutral: CLOSED
Earth (GND): OPEN
7. Read and record the current leakage indicated on the safety tester.
8. Configure the safety analyzer as follows (SFC):
Polarity: REVERSED
Neutral: CLOSED
Earth (GND): OPEN