4. In the Remote Service Configuration window, enter the applicable data:
5. Select Save.
The changes will take effect immediately.
7.16.2 Enabling remote service agent/ connection
After the server has been configured for remote serviceability, the remote service agent must
be enabled for use.
1. Log in to Webmin.
2. Select Configuration > Remote Service.
3. In the Sub-Modules for Remote Service menu, select Control.
4. In the Remote Service Control window, enable or disable the Service Agent by selecting
Enable or Disable.
5. Select Save.
The changes will take effect immediately.
HTTP Proxy Server Configuration
Item Description Comments
Address - If this site uses an HTTP proxy server, a
specific site proxy server IP Address and
Port number are required for the
Remote Service communication to
work. Otherwise, select None.
- If the HTTP proxy server requires user
authorization, a specific Username, and
Password is required. Otherwise, select
None.
These values are
determined by the
customer.
Port
Username and
Password
Remote Service Configuration
Item Description Comments
System ID Identifies the system to the GE back office
servers.
These values are read-only
and are unique.
Serial Number Identifies the unit and is set at the time of
manufacture.
Enterprise URL If required, designate the address of the GE
backoffice servers required to
communicate with the Remote Service
Agent.
This address should never
be changed unless explicit
instructions are given to do
so.
Enterprise Tunnel
URL
If required, designate the address of the GE
backoffice servers required to
communicate with the tunneling agent.
Protocol Identifies the protocol used to
communicate with the enterprise servers.
This field is read-only and
cannot be changed.

Table of Contents