CARESCAPE Monitor B850
76
2106778-001
Citrix Timeout: The number of minutes that a Citrix session will remain open after the
Citrix Window is closed. Keeping the session alive allows users to
quickly return to their last session but locks a Citrix license.
Citrix Startup Time: Estimated time, in seconds for connecting Citrix. During this time, the
Citrix Window is displayed as blank, grey rectangle.
7.7 Configuring MUSE/12SL
NOTE: MUSE viewing is a licensed software feature. You can configure the MUSE/12SL settings
independent of the license status, but actual viewing of MUSE reports requires that the
MuseView license is enabled.
Settings to send 12SL data
1. Log in to Webmin.
2. Select Configuration > MUSE/12SL.
3. In the MUSE/12SL window, enter the applicable information in the following fields:
4. Select Save.
These settings take effect immediately after they are submitted.
Settings to view 12SL data
1. Log in to Webmin.
2. Select Configuration > MUSE/12SL.
3. In the MUSE/12SL window, enter the following:
4. Select Save.
These settings take effect immediately after they are submitted.
- Location ID Identifies the location ID number (within the range 0 to 599)
associated with the patient monitor for searching the MUSE
system.
- Site Number Identifies the site number (within the range 1 to 254) associated
with the patient monitor for searching the MUSE system.
- MUSE Web Username Username used to authenticate with the MUSE
Web to access 12-lead reports.
- MUSE Web Password Password used to authenticate with the MUSE
Web to access 12-lead reports.
- Confirm Password The password and the confirmed password must
be the same.
- MUSE Web URL Used to locate the MUSE Web system to access
12-lead reports. Enter the URL in a valid format.