Troubleshooting
137
2106778-001
3. Select Download sGAS Logs.
4. According to your Webmin access:
a. If you are using Webmin on a service PC, the log file is saved to the default download
folder of the service PC.
b. If you are using Webmin locally through the integrated browser, you can save the log file
to a USB flash drive that is connected to one of the patient monitor's USB ports:
•The Download Logs window will show you the name of the created log file.
•Select Download to save the log file to the USB flash drive.
NOTE: Do not disconnect the USB storage device until downloading is completed.
5. Send the log file to GE Service for further investigation.
NOTE: The module is reset after you have downloaded the sGAS log. The reset takes up to 1
minute, as it calibrates the gas sensor and zeroes the measurement.
NOTE: To cancel the sGAS log download, you must admit a patient. When a patient is admitted,
the module is immediately reset. If you, for example, close the browser, the download
continues in the monitor and the module is reset only after that.
Viewing log files
1. Log in to Webmin.
2. Select Diagnostics > View Logs.
3. Select the log file you want to view.
4. Select the information you want to view.
For the Webmin Action Log, select the user, module, and timeframe and select
Search.
For the other types of logs, select the link associated with the information you want
to view.
Refer to the Module Frames and Modules Service Manual for information about PDM battery.
11.4 Error messages and codes
The following error messages display in the alarm area if there is a problem with the patient
monitor.
Refer to the patient monitor’s user’s manual for a complete list of system messages. Refer to
Module Frames and Modules Service Manual for a list of parameter module specific error
messages.
Message Possible causes Possible solutions
License expired A trial license has expired. Enable the license with a new
activation code, if needed.
Configuration
changes. Restart
required
Pending configuration
changes to platform settings
that require patient monitor
restart.
Power cycle the patient monitor.
Network down No other network device
observed on the MC Network.
Verify that the patient monitor is
connected to an active network.