1. Check that a network symbol is displayed in the upper right corner of the
screen.
2. Select Data & Pages – Other Patients and select a patient from the list.
3. Select View and verify that a window with parameters from another patient displays on
the left side of the screen.
4. Select Close View to close the window.
8.3.10 IX printers
Perform the following test only if the patient monitor is connected to a printer in the IX Network
and you did not print a test page while you configured the IX printer.
1. Log in to Webmin.
2. Select Configuration > Printers > Print Test Page.
3. Select the IX Printer to test.
4. Select Save.
5. Verify that the test page was printed to the selected printer.
6. Repeat steps 3 to 5 for all connected IX printers.
8.3.11 iPanel connection
Perform the following test only if the Citrix / iPanel is configured and in use.
1. Select Data & Pages > iPanel and verify that the initial program (configured in Webmin) is
launched correctly.
2. Select to exit the Citrix thin client.
8.3.12 InSite RSvP
Perform the following test only if remote service is configured and enabled, and you did not test
the connectivity already while you enabled it.
1. Log in to Webmin
2. Select Configuration > Remote Service.
3. In the Sub-Modules for Remote Service menu, select Control.
4. In the Remote Service Control window under Test Connectivity, select Test to test the
connectivity to the remote service server.
5. Verify that Connection to Enterprise URL and Connection to Tunnel URL both "Pass" the
test.
If one of the connection tests fail, check the failure description given and fix the connectivity