CARESCAPE Monitor B850
40
2106778-001
2. Continue to 4.4. Login to Webmin.
Closing Webmin
Select to close Webmin and return to the main display.
4.2 Local access to Webmin with a service PC
You can access Webmin locally by connecting an Ethernet crossover cable between the
service PC and the IX connector of the patient monitor.
NOTE: If you disconnect the patient monitor from a live IX Network when a patient is admitted,
you will temporarily lose the services provided by the IX Network, e.g., access to the IX printers
and MUSE reports.
WARNING Non-medical equipment does not provide the same level of protection
against electrical shock. Do not touch the patient and any part of
non-medical equipment at the same time. Some examples of non-medical
equipment are laser printers and non-medical computers.
1. Connect a service PC to the IX connector on the patient monitor using a crossover cable.
2. In patient monitor, select Monitor Setup > Service. The local browser opens and displays
the Login to Webmin dialog box.
3. Record the IX IP address of the patient monitor:
IX IP address: ______________________
IX Netmask: ______________________
NOTE: If the IX IP address field is shown as 0.0.0.0., you need to configure the patient monitor’s
IX Network address first. Access Webmin using the integrated browser and configure a static IP
address for the IX Network.
4. Configure the service PC’s IP address and subnet mask to the same network segment
with the patient monitor’s IX Network setting.