Hardware installation
59
2106778-001
6.3.3 Securing the USB cable connections of the displays
1. Secure the connected USB cables of all displays to the CPU unit. Use the existing retaining
clips attached to the patient monitor.
2. Secure the USB cables connections to the displays according to the instruction included
in the display package.
6.4 Installing module frames and modules
The patient monitor provides support to connect multiple parameter modules at a time. The
ePort/Tramnet ports allow connection of E-Module Frame and PDM.
See the following table for possible configuration options. For detailed information on
compatibility, refer to the the patient monitor’s supplemental information manual.
In the table above:
X = supported
Shaded = not supported
Example: If a PDM is connected to ePort 2, you cannot connect a second PDM to ePort 1.
However, you can connect one of the following devices to ePort 1:
•F5 Frame
•F7 Frame
WARNING ELECTRIC SHOCK - Do not use the F7 Frame for standalone use. Ventilation
holes on the F7 E-module Frame will be covered only if installed within an
Aisys CS2, Avance CS2, or Aespire anesthesia machine.
WARNING EQUIPMENT MALFUNCTION. Using the module frames in a wrong position
may result in equipment malfunction. The frames must only be used in
ePort 1 connected devices
PDM F5 Frame F7 Frame
ePort 2 connected devices PDM
XX
F5 Frame XX
F7 Frame X