CARESCAPE Monitor B850
164
2106778-001
FRU, CPU C1 Power supply
3. Reassemble the CPU C1 unit in the reverse order of disassembly.
12.2.9 Replacing the speaker
Disassemble first the following:
a. Opening the enclosure cover.
b. Removing the rear panel.
c. Remove the power supply according to Replacing the power supply
1. Disconnect the speaker cable from the CPU board.