Troubleshooting
145
2106778-001
11.5.3 Incorrect system time
11.5.4 License issues
Problem Possible cause Recommended action
System time is incorrect when
patient monitor is not
connected to network.
CPU timekeeper battery empty Replace CPU timekeeper battery.
Time not configured properly Configure date and time, see 7.2.4.
System time is incorrect when
patient monitor is connected to
MC Network.
Network device time
synchronization error.
When adding a new device to the
CARESCAPE Network, the existing devices
on the CARESCAPE Network will
synchronize to the new device’s time. To
prevent potential time synchronization
issues, you should set the new device’s
time to be as close as possible to the time
(within one minute or less) used by the
existing GE devices on the CARESCAPE
Network.
System time is incorrect when
patient monitor is connected to
S/5 network.
iCentral time incorrect. Configure iCentral time master.
Problem Possible cause Recommended action
Unable to perform a function or
a feature is not available.
- A license has not been
purchased for the feature.
- The trial license has expired
for the feature.
- The license is not installed
properly.
See License management chapter.