CARESCAPE Monitor B850
154
2106778-001
• GE recommends using the new fasteners (screws, washers, etc.) in the FRU kits rather
than reusing the old fasteners. Some fasteners are not intended to be re-used more than
three times.
• Use only new screws when attaching parts into light metal parts. Before fastening a
screw, turn it counterclockwise until it drops into an existing thread pattern.
12.1.3 Required tools
WARNING Due to possible high voltage present, use an insulated screwdriver at all
times.
12.1.4 Before disassembly
WARNING PATIENT MONITORING INTERRUPTION — Make sure a patient is not being
monitored while servicing the equipment.
WARNING DISCONNECTION FROM MAINS - When disconnecting the device from the
power line, remove the plug from the wall outlet first. Then you may
disconnect the power cord from the device. If you do not observe this
sequence, there is a risk of coming into contact with line voltage by
inserting metal objects, such as the pins of leadwires, into the sockets of the
power cord by mistake.
WARNING SAFETY GROUND - Remove power cord from the mains source by grasping
the plug. Do not pull on the cable.
WARNING ELECTRIC SHOCK - Always unplug the grounded cables when not in use.
Leaving them connected could result in an electric shock from the ground
contact in the other end.
1. Turn the monitor off from the rear On/off switch.
2. Detach the retaining clips for the power cord and USB cables.
NOTE: Remember to reattach the retaining clips for the power cord and USB cables after
reassembly.
- insulated Torx T10 screwdriver
- insulated crosshead screwdrivers (small and
medium)
- antistatic ESD wristband
- spanner size 5.5 mm