Installation checkout
113
2106778-001
NOTE: Perform this test for all the connected parameter modules and patient connectors
specified in Table 1.
1. Configure the safety analyzer as follows (NC):
Polarity: NORMAL
Neutral: CLOSED
Earth (GND): CLOSED
2. Power on the device under test.
3. Read and record the current leakage indicated on the safety tester.
4. Configure the safety analyzer as follows (SFC):
Polarity: NORMAL
Neutral: OPEN
Earth (GND): CLOSED
5. Read and record the current leakage indicated on the safety tester.
6. Configure the safety analyzer as follows (SFC):
Polarity: NORMAL
Neutral: CLOSED
Earth (GND): OPEN
7. Read and record the current leakage indicated on the safety tester.
8. Configure the safety analyzer as follows (SFC):
Polarity: REVERSED
Neutral: CLOSED
Earth (GND): OPEN
9. Read and record the current leakage indicated on the safety tester.
10. Configure the safety analyzer as follows (SFC):
Polarity: REVERSED
Neutral: OPEN
Earth (GND): CLOSED
11. Read and record the current leakage indicated on the safety tester.
12. Configure the safety analyzer as follows (NC):
Polarity: REVERSED
Neutral: CLOSED
Earth (GND): CLOSED
13. Read and record the current leakage indicated on the safety tester.
14. Power off the device under test.
15. Repeat this test for all the connected parameter modules and patient connectors
specified in Table 1.
Acceptance criteria in Normal Condition (NC):
• All readings shall be less than or equal to 10 A.
Acceptance criteria in Single Fault Condition (SFC) – earth open or one of the supply conductors
open at a time:
• All readings shall be less than or equal to 50 A.