Troubleshooting
139
2106778-001
Service Monitor
Error Code
0xHOST1100
The CPU timekeeper battery is
empty. Time and date may be
reset to factory settings.
Readjust the time and date first. Then
replace the CPU battery.
External alarm
light disconnect.
Check USB
connection.
1. The USB cable between the
patient monitor and a display
is disconnected when the
patient is admitted.
1. Reconnect the USB cable.
NOTE: Secure the connected USB cable
to the CPU unit. Use the existing
retaining clips attached to the CPU unit.
2. Display turned to standby
when the patient is admitted.
2. Turn on the display.
NOTE: Select Audio Pause to acknowledge the message, if the USB
cable is disconnected or a display turned off on purpose.
3. User interface board
communication failure.
3. UpdateUIC software or replace user
interface board.
Setting
activation after
next case end./
Setting
activation after
next discharge.
Service user has initiated a
delayed settings activation
that will automatically take
place after next case end. /
next discharge.
If required, the service user may cancel
the delayed setting activation. See
section 7.17.4. Canceling pending
settings activation
Software
activation after
next case end./
Software
activation after
next discharge.
Service user has initiated a
delayed software activation
that will automatically take
place after next case end /
next discharge.
If required, the service user may cancel
the delayed setting activation. See
section 7.20.3. Canceling pending host
software activation
Service Monitor
Activation Failed
Software activation failed. Reactivate software. If that does not
help, reinstall the software and then
reactivate it.
Setting activation failed. Reactivate settings.
Module
voltage low
• One of the supply voltages
for the acquisition
modules is out of
specification.
• An analog output voltage
is out of specification.
Log in to Webmin and select
Diagnostics > Hardware Stats to
diagnose the problem. If there is a
faulty board, replace it.
Application
error:
Webmin
The operating system informs
that the Webmin local
browser was terminated
abnormally.
Select Audio Pause to acknowledge
the message.
Message Possible causes Possible solutions