2. Remove the 8 screws that hold the processor board to the chassis. Remove the processor
board. Torque [1.1 Nm].
3. Remove the screw that holds the uDOM to the old processor board. Remove the uDOM
from the old processor board and reassemble it to the new processor board. Torque [0.3
Nm].