CARESCAPE Monitor B850
A - 2 (4)
2106778-001
Yes: Yes: you do not have to perform the electrical safety tests. To continue the
installation check, mark the electrical safety test completion N.A. and
proceed to Functional check. If available, attach the original electrical safety
test results to this check form.
No:
No: continue to perform the electrical safety tests.
Power outlet is correctly wired.
Power cord and plug are undamaged and all conductors are properly connected.
Ground integrity check Observed result Acceptance
criteria
PASS N.A. FAIL
a.) Ground continuity test without power cord 0.1 ohms
with power cord 0.2 ohms
b.) Impedance of protective earth
connection
without power cord 0.1 ohms
with power cord 0.2 ohms
Ground leakage current test Normal Condition (NC)
500 µA
EN /IEC
300 µA UL
Single Fault Condition (SFC) 1 mA
Testing touch current Normal Condition (NC) 100 µA
Single Fault Condition (SFC)
500 µA
EN /IEC
300 µA UL
Module type S/N Observed result Acceptance
criteria
PASS N.A. FAIL
ECG SpO2/
InvP
Patient (source) leakage current tests, using a test body
Normal Condition (NC) Polarity: NORMAL 10 µA
Single Fault Condition (SFC) Ground closed (normal) 50 µA
Ground open (normal) 50 µA
Ground closed (reverse) 50 µA
Ground open (reverse) 50 µA
Normal Condition (NC) Polarity: Reverse 10 µA
Patient (sink) leakage current tests
Polarity: NORMAL 50µA
Polarity: Reverse 50µA
Module type S/N Observed result Acceptance
criteria
PASS N.A. FAIL
Patient (source) leakage current tests, using a test body
Normal Condition (NC) Polarity: NORMAL 10 µA