EasyManua.ls Logo

GE CARESCAPE B850 - Page 50

GE CARESCAPE B850
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CARESCAPE Monitor B850
42
2106778-001
IX Netmask: ______________________
4. Configure the service PC’s IP address and subnet mask to the same network segment
with the patient monitor’s IX Network.
NOTE: For more information on how to configure the IP address, refer to the PC’s
documentation.
5. Launch a web browser on the service PC.
6. In the Address field, type https://[IX IP address]:10000 and press Enter.
NOTE: [IX IP address] is the IX Network IP address of the patient monitor.
The Login to Webmin dialog box displays.
7. Continue to 4.4. Login to Webmin.
Closing Webmin:
1. Restore the original network settings in the service PC.
2. Disconnect the service PC from the live IX Network.

Table of Contents