Using Webmin service interface
45
2106778-001
Webmin module Use the module
Admit Settings to configure Patient ID Prefix and barcode settings.
Certificate Management to manage SSL certificates installed on the system. With
this module, self-signed certificates and certificate
signing requests (CSR) can be generated. Certificates
can be imported, exported, deleted and viewed.
Citrix to configure the Citrix client settings for the iPanel
software: server address, initial program, Citrix session
timeout, username, password, encryption level.
Host Asset Settings to enter a host asset number and to view the host serial
number.
Language to select the language used in clinical user interface and
to select the keyboard locale setting for the
alphanumeric keyboard and the barcode reader.
Licenses to enable and activate a software package, to enable
and activate software features and to upload and
activate a license file.
MUSE/12SL to configure the host for sending and viewing 12SL
information.
Modules to configure some acquisition module specific settings
after corrective maintenance, or for administrative
purposes. These settings are saved to the permanent
memory of the related acquisition module and the
settings travel with the module from one patient monitor
to another.
Refer to the Module Frames and Modules Service
Manual for detailed information how to change these
settings.
National Requirements to activate France specific defaults for the ECG HR
adjustment range and the reminder beep behavior.
Network to select and configure the real-time network as the MC
Network (CARESCAPE Network or S/5 Network). It also
allows the user to configure the IX and ADT Network
settings.
Passwords to change the passwords for the biomed and clinical
users.
Power frequency to set the power line frequency.
Printers to configure the patient monitor to print to up to 12 laser
printers connected on the IX Network. There are
sub-modules for installing a printer, deleting a printer
and for printing a test page.
Remote Service to configure and control the InSite RSvP remote service
tool.

Table of Contents