2.4 Electrical Connection
WARNINGS:
1 VOLTAGES IN EXCESS OF 30 VOLTS (RMS) AC OR 50 VOLTS DC, IN
CERTAIN CIRCUMSTANCES, CAN BE LETHAL. CARE MUST BE
TAKEN WHEN WORKING ON LIVE, EXPOSED CONDUCTORS.
2 DO NOT DISCONNECT THE ADVANCED HAND TERMINAL WHEN
ENERGIZED IN THE HAZARDOUS AREA.
USE IN A HAZARDOUS AREA
SPECIAL CONDITIONS APPLY TO THE ATEX CERTIFIED ADVANCED HAND TERMINAL REFER TO PAGE V.
Power Supply
Single phase, 90 to 132 VAC, 47 to 400Hz, 200 VA.
180 to 260 VAC, 47 to 66Hz, 300 VA.
Power Supply Connection
The unit must be connected to the correct electrical power supply as stated, adjacent to the
power connector.
A qualified technician (see page i) must carry out the following procedure.
C
AUTIONS:
1T
HE SUPPLY MUST PROVIDE CONNECTION TO A PROTECTIVE GROUND TERMINAL. THE UNIT MUST, AT ALL
TIMES, BE CONNECTED TO THE SUPPLY EARTH (GROUND).
2T
HE POWER SUPPLY CABLE AND CONNECTOR MUST BE CORRECTLY RATED FOR THE POWER SUPPLY.
• Make sure that the power supply is off before connecting the power cable.
Fuses
The two fuses, located in the holders and mounted on the front panel, protect the unit. The fuses
are connected in the live and neutral supply circuit and are rated at:
5A anti-surge HBC 250V
External earth/ground connection
z An external earth (ground) cable may be connected to the stud on the front panel of the unit
providing integrity of the earth (ground) connection.
Advanced Hand Terminal Power Pack (Figure 2-3)
z Four interchangeable adaptors, supplied with the power pack, can be used to connect to the local
power supply. To change the power pin adaptor, push the adaptor from the power pack body,
align the replacement adaptor and push into the power pack body.
EUROPEAN
COLOUR
U.S.
COLOR
FUNCTION
Brown Black Live
Blue White Neutral
Green/Yellow Green
Protective Earth
(Ground)