Druck ADTS 2XX User Manual 6
K0451 Issue No. 1
2.5 Pneumatic Pressure Connections
Ps1 and Ps2 (static) - AN6 37° flare
Pt (pitot) - AN4 37° flare
When not in use, a blanking cap must be fitted.
Note: When carrying out a leak test, a leak of this blanking cap affects the performance of the ADTS 2XX.
Expansion ports
z Four sets of five expansion ports locate in the lid of the test set to provide hose connections for
pitot-static adaptors.
z When not in use, a blanking cap must be fitted to each connector.
FIGURE 2-1 ADTS 2XX LID-MOUNTED EXPANSION CONNECTIONS
Exp1
Ps1
S1 system (left)
S1 system (right)
Ps1 smart probe
not used - fit blanking cap
not used - fit blanking cap
Exp3
Ps2
S2 system (left)
S2 system (right)
Ps2 smart probe
not used - fit blanking cap
Exp2
Pt
P1 system (left)
P1 system (right)
Pt smart probe
Example Using Lid-mounted Expansion
Connection Ports