55 Druck ADTS 2XX User Manual
K0451 Issue No. 1
6 REFERENCE AND SPECIFICATION
6.1 General Specification
Weight ................................................................................................................................................ 22 kg [48.4 lb]
Hand Terminal ................................................................................................................................. 2 kg [4.8 lb]
Power Supply
Auto-selection between:
............................................................................................................... Single phase AC 90 to 260 V
.......................................................................................................... frequency range 47 to 440 Hz
Power ................................................................................................................................................................ 240 VA
EMC emmisions/immunity......................................................................................................................... EN61326-1
Safety performance
Electrical/mechanical......................................................................................................................................BS EN61010
Pressure equipment directive (PED), Class SEP, CE marked, ATEX Hand Terminal only, Zone 2
compliant
Shock and vibration
......................................................................................................... BS EN61010 MIL-T-28800 Class 2
Display
640 x 480 pixel, high contrast, backlit, colour LCD, resistive touch screen user interface
Environment
Temperature
The ADTS 505 has the following temperature ranges:
Calibrated ............................................................................................................. +5° to +35°C (+41° to +95°F)
Operating ............................................................................................................. 0° to +50°C (+32° to +122°F)
Storage ............................................................................................................. -20° to +70°C (-4° to +158°F)
Humidity
......................................................................................................................... 0 to 95% non condensing
Connections
Pressure connections
Ps1 and Ps2................................................................................................................................................... AN-6
Pt ............................................................................................................................................................... AN-4
Data communications
CANbus (hand terminal), RS232 (ext PC). All terminate on rugged 13-way circular MS38999 series
connector.