Druck ADTS 2XX User Manual 56
K0451 Issue No. 1
6.2 Air Data Specifications
Note: All range values at maximum limits.
Ps 1 Port
Pressure Mode
Control Range: ...................................................................................................................35
(1)
to 1355 mbar a
Measure Range: .................................................................................................................. 27.6 to 1355 mbar a
Resolution: ....................................................................................................................................................0.01mbar
Accuracy: .....................................................................................................................................................±0.1 mbar
Rate Range:............................................................................................................................0 to 1000 mbar/min
Rate Accuracy: .................................................................................................................. ±1% of aim (average)
Aeronautical Mode
Control Range: ........................................................................................................................ -2000 to 60,000 ft.
Measure Range: ..................................................................................................................... -3000 to 100,000ft.
Resolution: ................................................................................................................................................................ 1 ft.
Accuracy: ......................................................................................................................................±3 ft. (@sea level)
..............................................................................................................................................±7 ft. (@30,000 ft)
...........................................................................................................................................±23 ft. (@55,000 ft)
Rate Range:.................................................................................................................................0 to 40,000 ft/min
Rate Accuracy: .................................................................................................................. ±1% of aim (average)
Ps 2 Port
Pressure Mode
Control Range: ...................................................................................................................35
(1)
to 1355 mbar a
Measure Range: .................................................................................................................. 27.6 to 1355 mbar a
Resolution: ....................................................................................................................................................0.01mbar
Accuracy: .....................................................................................................................................................±0.1 mbar
Rate Range:............................................................................................................................0 to 1000 mbar/min
Rate Accuracy: .................................................................................................................. ±1% of aim (average)
Aeronautical Mode
Control Range: ........................................................................................................................ -2000 to 60,000 ft.
Measure Range: ..................................................................................................................... -3000 to 100,000ft.
Resolution: ................................................................................................................................................................ 1 ft.
Accuracy: ......................................................................................................................................±3 ft. (@sea level)
..............................................................................................................................................±7 ft. (@30,000 ft)
...........................................................................................................................................±23 ft. (@55,000 ft)
Rate Range:.................................................................................................................................0 to 40,000 ft/min
Rate Accuracy: .................................................................................................................. ±1% of aim (average)
(1)
limited only by vacuum pump/system leaks