43 Druck ADTS 2XX User Manual
K0451 Issue No. 1
4.4 Routine Maintenance
Absolute cleanliness of the work area, tools and equipment are essential.
Replacing the output connector o-ring
After inspection as detailed in maintenance task B, carry out the
following if the o-ring is worn or damaged:
Carefully remove the o-ring from the small groove at the top of the
connector. Fit a new o-ring in the small groove at the top of the
connector.
Make sure the o-ring is tight in the groove and not damaged after
fitting.
Note: Damage to this o-ring causes leaks.
Replacing a fuse
Disconnect the electrical power supply.
Unscrew the fuse holder cap and remove the fuse.
Fit a new fuse of the correct type and rating into the fuse holder
cap.
Secure the fuse holder in the front panel.
Repair of the AHT
The AHT is a non-serviceable component. If the AHT becomes unserviceable, it can only be
replaced by another ATEX compliant AHT.