Druck ADTS 2XX User Manual 30
K0451 Issue No. 1
STAT
This group of commands monitor and control the status bytes that show the status of the ADTS 2XX.
STAT:OPER:CON
Command Syntax
N/A
Parameter Type:
Parameter Range:
Function:
Conditions:
Query Syntax
STATus:OPERation:CONdition?
Parameter Type: None
Parameter Range:
Returned Data: <status>
Returned Data Type: INTEGER
Returned Data Range: 0-32767
Function: To query the status of the ADTS. The table shows the function of each bit.
Conditions: None
Bit Data
0 Reserved – returns 0
1 Ps1 controlling and at set-point
2 At ground
3 Ps2 controlling and at set-point
4 Pt controlling and at set-point
5 Reserved – returns 0
6 Reserved – returns 0
7 Reserved – returns 0
8 Reserved – returns 0
9 Reserved – returns 0
10 Reserved – returns 0
11 Reserved – returns 0
12 Reserved – returns 0
13 Reserved – returns 0
14 Reserved – returns 0
15 Reserved – returns 0