51 Druck ADTS 2XX User Manual
K0451 Issue No. 1
TABLE 5-2 ERROR MESSAGES (CONTINUED)
245 "Controller has an invalid EEPROM
archive"
Invalid/corrupted data - recovery
in progress
Check/correct calibration
247 "Handset has an invalid EEPROM
archive"
Invalid/corrupted data - recovery
in progress
250 "Vacuum pump processor has
failed"
Internal fault Return unit to repair depot
251 "Vacuum pump lead is
disconnected"
- Return unit to repair depot
252 "Vacuum pump transistor is faulty
or drive inhibited"
Internal fault Return unit to repair depot
No. Message Probable Cause Action
253 "Vacuum pump has stalled" Internal mechanical fault Restart system, if fault repeats -
return unit to repair depot
254 "Pressure pump processor has
failed"
Internal fault Return unit to repair depot
255 "Pressure pump lead is
disconnected"
- Return unit to repair depot
256 Pressure pump transistor is faulty
or drive inhibited"
Internal fault Return unit to repair depot
257 Pressure pump has stalled" Internal mechanical fault Restart system, if fault repeats -
return unit to repair depot
260 Connected handset has failed to
respond. Disconnecting power."
Handset bad connection or
comms/power supply fault
Check connections and restart, if
fault repeats - return unit to
repair depot