v Druck ADTS 2XX User Manual
K0451 Issue No. 1
ATEX CERTIFIED ADVANCED HAND TERMINAL
CONDITIONS OF USE
The ATEX certified Advanced Hand Terminal can be used in zone 2* hazardous areas in accordance with the
ATEX certification document and schedule.
Marking detail:
Refer to Advanced Hand Terminal User Manual K0418 and the label on the Advanced Hand Terminal.
SPECIAL CONDITION OF USE
The power supplies must be isolated when connecting the advanced hand terminal in the hazardous
area.
The advanced hand terminal must not be disconnected when energized in the hazardous area.
The advanced hand terminal is a non-serviceable component. If the advanced hand terminal
becomes unserviceable it can only be replaced by another ATEX compliant hand terminal.
Note:
The advanced hand terminal must only be used with the cable assembly supplied and marked “DO NOT
SEPARATE WHILST ENERGISED IN HAZARDOUS AREA
* Zone 2 hazardous area definition, see page 53.