13.3 Manufacturer’s Guidance and Declaration Regarding
Electromagnetic Immunity
The Girae Blue Spot PT Lite is intended for use in the electromagnetic environment specied below. The
customer or the user of the Girae Blue Spot PT Lite should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment
Guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
+/- 6kV contact
+/- 8kV air
+/- 6kV contact
+/- 8kV air
Floors should be wood, concrete or
ceramic tile. If oors are covered
with synthetic material the relative
humidity should be at least 30%.
Electrical fast transient/
burst IEC 61000-4-4
+/- 2kV for power supply
line.
+/- 2kV for power
supply line.
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC 61000-4-5 +/- 1kV dierential Mode.
+/- 2kV common mode
line.
+/- 1kV dierential
Mode.
+/- 2kV common
mode line.
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, shorts
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
<5 % Ut (>95 % dip in Ut)
0.5 cycle
40 %Ut (60 % dip in Ut)
for 5 cycles
70 % Ut (30 % dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
<5 % Ut (>95 % dip in
Ut) 0.5 cycle
40 %Ut (60 % dip in
Ut) for 5 cycles
70 % Ut (30 % dip in
Ut) for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 sec.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Girae Blue Spot PT Lite requires
continued operation during power
mains interruption, it is recommended
that the Girae Blue Spot PT Lite be
powered from an uninterruptible
power supply.
Power frequency
(50/60 Hz) magnetic
eld environment
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds
should be at levels characteristic
of a typical location in a typical
commercial or hospital.
NOTE: Ut is the AC main voltage prior to application of the test level.
91
Chapter 13: Electromagnetic Guidance and Declarations