13.3 Manufacturer’s Guidance and Declaration Regarding
Electromagnetic Immunity
The Girae Blue Spot PT Lite is intended for use in the electromagnetic environment specied below. The
customer or the user of the Girae Blue Spot PT Lite should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment
Guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
+/- 6kV contact
+/- 8kV air
+/- 6kV contact
+/- 8kV air
Floors should be wood, concrete or
ceramic tile. If oors are covered
with synthetic material the relative
humidity should be at least 30%.
Electrical fast transient/
burst IEC 61000-4-4
+/- 2kV for power supply
line.
+/- 2kV for power
supply line.
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC 61000-4-5 +/- 1kV dierential Mode.
+/- 2kV common mode
line.
+/- 1kV dierential
Mode.
+/- 2kV common
mode line.
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, shorts
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
<5 % Ut (>95 % dip in Ut)
0.5 cycle
40 %Ut (60 % dip in Ut)
for 5 cycles
70 % Ut (30 % dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
<5 % Ut (>95 % dip in
Ut) 0.5 cycle
40 %Ut (60 % dip in
Ut) for 5 cycles
70 % Ut (30 % dip in
Ut) for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 sec.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Girae Blue Spot PT Lite requires
continued operation during power
mains interruption, it is recommended
that the Girae Blue Spot PT Lite be
powered from an uninterruptible
power supply.
Power frequency
(50/60 Hz) magnetic
eld environment
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds
should be at levels characteristic
of a typical location in a typical
commercial or hospital.
NOTE: Ut is the AC main voltage prior to application of the test level.
91
Chapter 13: Electromagnetic Guidance and Declarations