13.4 International Electronic Commission (IEC) Guidance and
Manufacturer’s Declaration Regarding Electronic Immunity
The Girae Blue Spot PT Lite is intended for use in the electromagnetic environment specied below. The
customer or the user of the Girae Blue Spot PT Lite should assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance Level Electromagnetic Environment Guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the Girae Blue Spot PT Lite, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended Separation Distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 KHz to 80 MHz
3V
d =1.2√ P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
IEC 60601-2-19 &
IEC 60601-2-21
3 V/m
26 MHz to 1 GHz
3 V/m
normal operation
d =1.2√ P = 26 MHz to 800 MHz
10 V/m
26 MHz to 1 GHz
10 V/m
no hazard
d =2.3√ P = 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths for xed RF transmitters
as determined by an electromagnetic site
survey (see Note 3-a) should be less than the
compliance level in each frequency range (see
Note 3-b).
Interference may occur in the vicinity of
equipment. Marked with the following symbol:
NOTE 1: Portable and mobile RF equipment can aect medical electronic equipment.
NOTE 2: At 80 MHz and 800 MHz, the higher frequency applies . These guidelines may not apply in all