IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
8-28 Chapter 8 - Replacement Procedures
PRELIMINARY
8-2-10 C-FEB Replacement Procedure
8-2-10-1 Tools
Phillips screwdriver of appropriate size
8-2-10-2 Time Required
5 min
8-2-10-3 Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in Power Shutdown.
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Go™ system from cart as described in Mounting / Dismounting System on Cart
and place it on a previously prepared surface with the display facing down.
8-2-10-4 C-FEB Removal Procedure
1) Remove back cover.
2.) Remove the C-PSB.
3) Remove the fan.
4.) Remove the C-BEB.
5.) Disconnect the front display indicator cable (marked with blue circle).
FRU Part #
Refer to Table 9-16 on page 9-17.
Back Cover Removal Procedure
C-PSB Removal Procedure
Fan Removal Procedure
C-BEB Removal Procedure