IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
8-42 Chapter 8 - Replacement Procedures
PRELIMINARY
8-2-16 Cradle Tilt Replacement Procedure
8-2-16-1 Tools
4mm Allen key.
8-2-16-2 Time Required
5 min
8-2-16-3 Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in Power Shutdown .
2.) Prepare stable flat surface covered with a soft material (such as foam) to prevent scratches to the
display.
3.) Dismount Venue Go™ system from cart and place it on a previously prepared surface with the
display facing down.
8-2-16-4 Cradle Tilt Removal Procedure
4.) Remove the cradle assembly.
5) Raise the cart to its highest position.
6.) Slide plastic cap upwards and remove it
FRU Part #
Refer to Table 9-16 on page 9-17.
Cradle Assembly Removal Procedure
Figure 8-121 Removing Column Cover

Table of Contents