IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-53
PRELIMINARY
8-2-21 Cart Up/Down Handle Replacement Procedure
8-2-21-1 Tools
Phillips screwdriver of appropriate size
8-2-21-2 Time Required
5 min
8-2-21-3 Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in Power Shutdown .
2.) Prepare stable flat surface covered with a soft material (such as foam) to prevent scratches to the
display.
3.) Dismount Venue Go™ system from cart and place it on a previously prepared surface with the
display facing down.
8-2-21-4 Up/Down Handle Removal Procedure
4.) If ECG is installed in the up/down handle’s screws side, remove the ECG
5.) Remove the 3 screws holding the Up/Down handle. .
8-2-21-5 Up/Down Handle Installation Procedure
1) Install the Up/Down Handle on the cart using 3 screws.
2.) If removed, re-install ECG assembly.
FRU Part #
Refer to Table 9-16 on page 9-17.
ECG Removal Procedure
Figure 8-133 Disconnect Up/Down Handle Screws
ECG Installation Procedure