IRECTION 5813707-100, REVISION 2 VENUE GO™ SERVICE MANUAL
10-10 Chapter 10 - Care and Maintenance
PRELIMINARY
10-5-4 Cleaning
10-5-4-1 General Cleaning
Frequent and diligent cleaning of the Venue Go™ ultrasound unit reduces the risk of spreading infection
from person to person, and also helps to maintain a clean working environment.
Use only cleaning materials and solutions as recommended in the procedures described in the
Venue Go™ User Manual.
Do not use any solutions or products not listed in the Venue Go™ User Manual.
Never use thinner, benzene, ethanol or methanol alcohol, abrasive cleaners, or other strong
solvents. Only use isopropyl alcohol, when instructed to do so.
Do not spray any liquid directly onto the Venue Go™ covers.
Do not allow any liquid to drip or seep into the system.
DO NOT scratch or press on the panel with any sharp objects, such as pencils or pens, as this may
result in damage to the panel.
Make sure not to spill or spray any liquid on the controls, into the Venue Go™ cabinet, or in the
probe connection receptacle.
Prior to cleaning, turn OFF power to the Venue Go™ and disconnect the mains cable.
NOTE: Refer to the Venue Go™ User Manual for cleaning instructions
CAUTION
WHEN PERFORMING CLEANING PROCEDURES, TO PREVENT THE RISK OF SYSTEM
DAMAGE, ALWAYS OBSERVE THE FOLLOWING PRECAUTIONS: