D
IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-35
PRELIMINARY
3.) Connect the ground cable: Attach the grounding-cable’s rings and the cable clamp to the dedicated
Assy, as shown in the picture. Make sure the ring lugs are physically touching the chassis.
4.) Secure the two cable clamps using two screws. Fasten the screws by 0.62 torque.
5.) Connect the speakers’ cable to the C-BEB.
6.) Install the fan.
7.) Install back cover.
Figure 8-113 Speakers Earth Cable
• Fan Installation Procedure
• Back Cover Installation Procedure