8-16 Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-3-8 Modify System/Windows Settings
After Base System is successfully installed, various parameters must be properly set before proceeding to Application
load. The below outlines necessary system/windows setting.
Table 8-66 Summery of modification
Item Method Proper Value
UPS Setting Automatic Setting (n/a – just load)
(Restart System)
Avance AC97 Audio Load from c:\windows\system32\drivers (n/a – just load)
(Restart System)
Drive Letter Automatic Setting but check desktop icon if not. DVD Drive = G
Matrox for CRT Check off automatic pop-up window Set Taskbar into CRT Display
CRT Display mode: 800 x 600 75Hz
Touch Panel Display mode: 640 x
480 60Hz
HW Device My Computer
\
Properties
Various
LAN Desktop
\
My Network Places
\
Properties
\
Local Area Connection Double Click to activate
TCP/IP Setting Desktop
\
My Network Places
\
Properties
\
Local Area Connection
\
Internet Protocol (TCP/IP)
\
Properties
\
Advanced
\
Options
\
TCP/IP Filter Values in ‘Permit Only’
Date/Time Start
\
Settings
\
Control Panel
\
Date and Time
Proper Time Zone
Accurate Date / Time