4-4 Transport Pro Revision A
2012659-042
Configuration: Monitor Settings
Unit Name
The unit name identifies the care unit (up to seven characters). The unit
name displays at the top right of the display before the bed number.
To define the unit name, follow this procedure.
1. Select MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Use the TRIM KNOB control to enter the service password.
3. Select MONITOR SETTINGS > UNIT NAME.
4. Use the TRIM KNOB control to select up to seven characters.
5. Select SET UNIT NAME and press the TRIM KNOB control to exit.
Bed Number
The bed number identifies a particular patient bed (up to five
characters). The bed number displays at the top right of the display after
the unit name.
To define the bed number, follow this procedure.
1. Select MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Use the TRIM KNOB control to enter the service password.
3. Select MONITOR SETTINGS > SET BED NUMBER.
4. Use the TRIM KNOB control to select up to five characters.
5. Select SET BED NUMBER and press the TRIM KNOB control to
exit.
Country Selection
The country selection defines a particular set of GE factory defaults.
There are two options: DEFAULT or FRANCE.
To select the country, follow this procedure.
1. Access the Boot Loader program.
2. Select SET CONFIGURATION > Country Selection.
↓ ↑