Troubleshooting: Problems
Missing/Inaccurate Patient Data/Waveforms
1. Refer to operator’s manual for detailed information on patient
preparation and electrode placement.
2. Change the sensitivity level. Usually a lower percentage rectifies the
problem.
3. Test the patient cable on a known working device. If the patient cable
is defective, replace.
4. Test the patient connector with a patient simulator. If the patient
connector is defective, replace the acquisition device.
Speaker Buzz
During Alarms
1. Reboot the monitor.
2. Reseat speaker cable connection.
3. Verify that speaker and/or cable is not damaged.
4. If speaker or cables are damaged, replace the speaker FRU.
5. Unplug speaker and check speaker cable resistance.
6. If resistance is between 6 and 8 ohms, replace the main processor
FRU.
7. If resistance is outside 6 and 8 ohms, replace the speaker.
Always
1. Reboot the monitor.
2. Replace the main processor FRU.
Unresponsive Buttons
1. Reboot the monitor.
2. Connect a known good power supply.
3. Reseat the keypad harness connection.
4. Verify that the keypad harness is not damaged.
5. If the keypad harness is damaged, replace the front bezel.
6. If the keypad harness is not damaged, replace the main processor
FRU.