Chapter 7. Configuring the Transmitter
120 TransPort
®
PT900 User’s Manual
7.3.2c Programming the PEAK DETECT Option
From the drop-down list in the PEAK DETECT option, select the desired method for
identifying the peak of the received signal. The following options are available:
The PEAK method is no longer available.
With the THRESHOLD method, the peak is identified as the point where the
signal crosses a threshold that is a percentage of the maximum signal
detected. This method is more reliable in marginal signal conditions.
7.3.2d Programming the THRESHOLD Option
If the THRESHOLD method is selected in the PEAK DETECT option, the THRESHOLD
DETECT
parameter must be configured. From the drop-down list, select one of the
following options:
With the Auto method, the threshold is detected automatically.
With the Manual method, the minimum and maximum threshold
percentage (0 to 100%) must be input. The
Peak Percentage is also
required.
7.3.2e Saving Your Settings
Important: Be sure to click the SAVE button before leaving the METER SETUP
menu, or all of your settings will be discarded.