139 GSM19 Series Magnetometers - Version 7
GEM Systems, Inc. Advanced Magnetometers
For more technical information, visit www.gemsys.ca
APPENDIX I: PROGRAMMABLE BASE OPTION
The programmable base feature allows full control over the base surveys. It is particularly useful for
remote or unattended base stations.
By programming or scheduling the base, you can determine when the surveys begin and end hours,
days or years in advance. This result in a more efficient use of batteries or other available power
sources, and also avoids storing in memory hours or days-worth of unwanted data. Because a new
file is created every time a base survey starts, each schedule has its own file.
I.1 SETTING UP THE PROGRAMMABLE BASE STATION OPTION
From the Main Menu, press A-survey to access the Survey Menu and then, with survey mode
highlighted, press
C-change.
NOTE:
If GPS synchronization is required make sure the GPS is enabled.
SCREEN G1
A - mobile B - base
C - grad
D - walkmag E - walkgrad
NOTE:
The available survey modes of your unit may differ from the ones shown here
depending on its model, its software, and its hardware.
Press B
to select Base Station. The unit returns to Survey Menu with the following screen:
SCREEN G2
survey mode datum time file
cycle time tuning AC filter
display mode text ID
connect sensor now
089123 readings left
base
A-start C-change
←
BF
→