G 2 00 R A N G E
EVENT LOG
The following events are recorded in the instrument’s event log. The event log can only be
downloaded using the additional Analyser Data Manager software. Please refer to the Analyser Data
Manager operating manual for further details.
Event
Data
Cold Start/Reset
Type of start (MCUSR, boot_key)
Firmware version
Description
Set time
Before and After
Set date
Before and After
Re-flash requested
None
Restore to factory settings
Type
Comms clear memory
Type 0=Readings, 1=Event Log
Change contrast
Before and After
Change flow fail current limit
Before and After
RTC date/time invalid
None
Factory calibration invalid or overdue
Date, Difference
Service invalid or overdue
Date, Difference
Field/User Calibration invalid or overdue
Time, Channel, Date, Age, Limit
O2 Cell replacement overdue
Time, Channel, Date, Age, Limit
Pump replacement overdue
Used, Limit
Battery replacement overdue
Used, Limit
EEProm settings set to default
Which
Missing EEProm settings initialised
Which, Orig Version, New Version
CO cell replacement overdue
Time, Channel, Date, Age, Limit
Power on self-test, sensor not found
Channel
Power in self-test, sensor out of range
Channel, Reading, Low, High
Power on self-test, sensor out of range
Channel, Reading, Prev
Battery dropped below critical value
Limit, Actual
User calibration set zero OK
Before and After
User calibration set span OK
Before, After
User calibration set zero failed
Target, Reading
User calibration set span failed
Target, Reading
Attempt to store when readings memory full
Max
Readings memory nearly full
Limit, Actual
Flow fail current limit exceeded
Limit, Actual
Automatic power off
Change logging mode
Status, Interval, Pump time
Change logging mode ID
ID
Changing logging mode ID site
Site ID
Flow fail current limit exceeded
Limit, Actual
An alarm has changed state
Alarm, State, Actual, High, Low
Change of alarm setting
Alarm, State, High, Low
Mode of instrument changed
Before, After