5 DETERGENT
5.1 Using detergent
WARNING!
Dosing monitoring only works with detergents
with viscosities lower than 10 cSt.
WARNING!
Do not use detergent for household dishwashers.
Note!
The specified performance may not be achieved if other
process chemicals than those which were tested during
type testing are used.
The machine has been proven and tested with Getinge's range of
detergents as per ISO 15883-1:2009.
If chemicals are recommended in the manual, this does not
mean that the manufacturer of the machine accepts liability
for any impact the agent has on goods to be washed.
Changes in the composition, conditions, etc. that were not
reported by the manufacturer may affect the result of the
cleaning process.
Always follow the instrument manufacturer's instructions
when you use new instruments.
The dosage amount depends on the water quality. For more
information, see the process chemicals product
specifications.
5 DETERGENT
INSTALLATION MANUAL GETINGE WD15 CLARO 35 (48)
<Doc_INS><Doc_6001396702><Rev.I><Lang_en>

Table of Contents

Other manuals for Getinge GE15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge GE15 and is the answer not in the manual?

Getinge GE15 Specifications

General IconGeneral
BrandGetinge
ModelGE15
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals