Safety-related information
Safety instructions
2
LUCEA 50-100
IFU 01741 EN 11
25 / 60
2.2.2 Electrical
CAUTION!
Risk of malfunction of the device
The use of accessories, transducers or cables other than those supplied or
recommended by the manufacturer of this device may cause increased elec-
tromagnetic emissions or a decreased immunity of this device, and may res-
ult in improper operation.
Use only accessories and cables supplied or specified by the manufacturer.
WARNING!
Risk of electric shock
Anyone not trained in installation, maintenance or decommissioning opera-
tions is exposed to the risk of injury or electric shock.
Installation, maintenance and decommissioning of the device or components
of the device must be performed by a Getinge technician or a Getinge-trained
service technician.
WARNING!
Risk of injury
If a power cut occurs in the middle of an operation, the lightheads will go out
if the lighting system does not have a backup supply.
The hospital must comply with applicable standards on premises for medical
use and must have a backup power system.
2.2.3 Optical
WARNING!
Risk of burns
The high intensity of the light source results in a risk of burns to the eye if
the user looks directly towards the lighthead.
The patient’s eyes must be protected during facial surgery. Users must not
look directly into the light source.
2.2.4 Infection
WARNING!
Risk of infection
A maintenance or cleaning operation may result in contamination of the sur-
gical site.
Do not perform maintenance or cleaning operations when the patient is
present.