4
Use
Mobile light
40 / 60
LUCEA 50-100
IFU 01741 EN 11
4.6.2 Battery system operation
1
2
3
4
Fig.32: Battery system indicators
Operation when the mobile light is connected to the mains
When operating on mains power, the LED with the plug icon 1 is green.
While the batteries are charging, LEDs3 to8 3 scroll.
Once the batteries are fully charged, LED8 4 starts flashing.
NOTE
It takes at least 10hours to fully charge the batteries.
Operation when the mobile light is on battery power
When operating on battery power, the LED with the battery icon 2 is green.
If a power cut occurs, the lighting runs on battery power, and the batteries thus gradually dis-
charge.
The charge in the batteries is indicated by LEDs3 to8 3 . As the batteries run down, the in-
dicator moves from (+) towards (-).
When the batteries are discharged, an alarm signal sounds and LED2 2 turns red.
The light shuts off automatically after the alarm signal sounds (to protect the batteries against
deep discharge).
NOTE
If the batteries are fully charged, the LUCEA 50 can run on battery power for at
least three hours and the LUCEA 100 for eight hours.

Table of Contents

Other manuals for Getinge Lucea 100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Lucea 100 and is the answer not in the manual?

Getinge Lucea 100 Specifications

General IconGeneral
BrandGetinge
ModelLucea 100
CategoryMedical Equipment
LanguageEnglish

Related product manuals