Cleaning / Disinfection / Sterilisation
Cleaning and disinfecting the system
7
LUCEA 50-100
IFU 01741 EN 11
47 / 60
7.1.2 Disinfecting the device
Wipe evenly with a cloth soaked in disinfectant. Follow the manufacturer’s recommendations.
7.1.2.1 Disinfectants to be used
Disinfectants are not sterilising agents. They result in a qualitative and quantitative reduction
in the microorganisms present.
Use only surface disinfectants containing combinations of the following active substances:
Quaternary ammoniums (bacteriostatic for Gram– and bactericidal for Gram+, variable
activity on enveloped viruses, no action on non-enveloped viruses, fungistatic, no spori-
cidal action)
Guanidine compounds
Alcohols
7.1.2.2 Permitted active substances
Class Active substances
Low level of disinfection
Quaternary ammonium Didecyl dimethyl ammonium chloride
Alkyl dimethyl benzyl ammonium chloride
Dioctyl dimethyl ammonium chloride
Biguanides Polyhexamethylene biguanide hydrochloride
Intermediate level of disinfection
Alcohols Propan-2-ol
High level of disinfection
Acids Sulfamic acid (5%)
Malic acid (10%)
Ethylene diamine tetraacetic acid (2.5%)
Tab.14: Lists of active substances suitable for use
Examples of commercially available products tested
ANIOS product®** : Surfa’Safe®**
Other products: 20% or 45% isopropyl alcohol

Table of Contents

Other manuals for Getinge Lucea 100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Lucea 100 and is the answer not in the manual?

Getinge Lucea 100 Specifications

General IconGeneral
BrandGetinge
ModelLucea 100
CategoryMedical Equipment
LanguageEnglish

Related product manuals