Inspections
Electrical inspections
6
Maquet PowerLEDII
01810 EN 03
37 / 56
Fig.49: Wall-mounted power supply
• Check the tightening of the connections.
• Check the ground connections.
• Check the general condition of the power
supply.
Power Supply
XXXXXXXXXX
Power Supply
XXXXXXXXXX
Fig.50: Backup
• Check the tightening of the connections.
• Check the ground connections.
• Verify the output voltage of the batteries
(12 V per battery, difference of voltage
between batteries <0.1 V).
• Check the general condition of the backup,
the batteries, the absence of swelling,
leaks, oxidation.
6.2.2 Performing battery tests
WARNING!
Risk of injury
A battery lifetime test fully discharges the batteries.
Do not perform an operation immediately after a battery lifetime test. Allow
time for the batteries to charge.