6
Inspections
Electrical inspections
38 / 56
Maquet PowerLEDII
01810 EN 03
6.2.2.1 From the wall-mounted control keypad
1
2
3
Fig.51: Battery tests from the wall-mounted keypad
Running a battery backup test
1. Turn off the light.
2. Press Battery backup test 1 .
Ø If the test is successful, the battery level indicator 2 flashes green. If the test fails, the
battery level indicator 2 flashes red.
3. If the test fails, contact the Getinge technical service department.
4. Press Battery backup test 1 again.
Ø The battery level indicator 2 stops flashing. The light is on and ready for use.
Running a battery lifetime test (only with a Getinge backup)
1. Turn off the light.
2. Press Battery lifetime test 3 .
Ø If the test is successful, the battery level indicator 2 flashes green. If the test fails, the
battery level indicator 2 flashes red.
3. If the test fails, contact the Getinge technical service department.
Ø The light turns off when the test is complete.
4. Press Battery lifetime test 3 again.
Ø The battery level indicator 2 stops flashing.
NOTE
The battery lifetime test can be stopped at any time by pressing Battery lifetime
test 3 again until the lightheads turn off.

Other manuals for Getinge Maquet PowerLED II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet PowerLED II and is the answer not in the manual?

Getinge Maquet PowerLED II Specifications

General IconGeneral
BrandGetinge
ModelMaquet PowerLED II
CategoryMedical Equipment
LanguageEnglish

Related product manuals