P Parameter Default
value
Lower
limit
Upper
limit
Unit Description
CT10 W INT 1 DOS
PUMP
0 0 4 - Dosing pump Intensive Wash 1
0=No dosing
1=Pump 1
2=Pump 2
3=Pump 3
4=Pump 4
CT11 W INT 2 DOS
PUMP
0 0 4 - Dosing pump Intensive Wash
dose 2
0=No dosing
1=Pump 1
2=Pump 2
3=Pump 3
4=Pump 4
CT12 NEUT DOS PUMP 0 0 4 - Dosing pump Neutralization.
0=No dosing
1=Pump 1
2=Pump 2
3=Pump 3
4=Pump 4
CT13 F RINSE DOS
PUMP
2 0 4 - Dosing pump Final rinse
0=No dosing
1=Pump 1
2=Pump 2
3=Pump 3
4=Pump 4
CT15 WATER CONS #1 12 0 12 - Water quality (only for
machines with salt filter)
LI08 WASHING DOS
TEMP
35.0 20.0 85.0 C Dosing temperature washing
LI09 WASH INT D1 TMP 35.0 20.0 85.0 C Dosing temperature intensive
washing step 1
LI10 WASH INT D2 TMP 55.0 20.0 85.0 C Dosing temperature intensive
washing step 2
LI11 NEUTR DOS TEMP 40.0 20.0 85.0 C Dosing temperature
Neutralization
LI12 F RINSE DOS TEMP 80.0 20.0 85.0 C Dosing temperature Final rinse
LI14 A0_VALUE
60.0 600 9000 F0 A0 value
LI15 COOL TEMP
CHAMB
80 50 80 C Temperature cooling
LI67 DRAIN COOLING 60.0 40.0 75.0 C Temperature limit for drain
cooling
LI68 CONDENSER CNTR 60.0 40.0 75.0 C Temperature limit for
condensate cooling
3 MENU TREE
46 (128) TECHNICAL MANUAL GETINGE WD15 CLARO
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>