P Parameter Default
value
Lower
limit
Upper
limit
Unit Description
PF00 PRE RINSE 1 SEL 1 0 1 - Activate Pre-rinse 1
0=Off
1=On
PF01 PRE RINSE 2 SEL 1 0 1 - Activate Pre-rinse 1
0=Off
1=On
PF02 WASHING SEL 1 0 1 - Activate Wash
0=Off
1=On
PF03 WASH INT SEL 0 0 1 - Activate Intensive wash
0=Off
1=On
PF04 NEUTR SEL 0 0 1 - Activate Neutralization
0=Off
1=On
PF05 RINSE 1 SEL 1 0 1 - Activate Post-rinse 1
0=Off
1=On
PF06 RINSE 2 SEL 1 0 1 - Activate Post-rinse 2
0=Off
1=On
PF07 RINSE 3 SEL 0 0 1 - Activate Post-rinse 3
0=Off
1=On
PF08 FINAL RINSE SEL 1 0 1 - Activate Final rinse
0=Off
1=On
PF09 DRYING SEL 1 0 1 - Activate Drying
0=Off
1=On
PF27 INTENSIV WASH 0 0 1 C Activate Wash step Intensive
0=Off
1=On
PF28 DISINFECTION
0 0 1 C Activate Disinfection step
0=Off
1=On
PF31 START
PASSWORD
0 0 1 - Activate password-protected
start for a program.
0=Off
1=On
PF34 A0-PRINT 0 0 1 C Activate printout of A0 value
reached
0=Off
1=On
3 MENU TREE
TECHNICAL MANUAL GETINGE WD15 CLARO 47 (128)
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>